Abstract: Hybrid retinal imager using line-scanning laser ophthalmoscopy and spectral domain optical coherence tomography

Nicusor V. Iftimia, Daniel X. Hammer, Chad E. Bigelow, Teoman Ustun, Johannes F. de Boer, R. Daniel Ferguson, "Hybrid retinal imager using line-scanning laser ophthalmoscopy and spectral domain optical coherence tomography ," Optics Express 14 (26) , 12909-12914 (25 December2006).

Article: 1566 kB

Reprinted by permission of the Optical Society of America. For permission to copy or republish, contact the Optical Society of America, 2010 Massachusetts Ave. NW, Washington, DC 20036-1023.

Abstract

We demonstrate for the first time the integration of two technologies, Spectral Domain Optical Coherence Tomography (SDOCT) and Line-Scanning Laser Ophthalmoscopy (LSLO) into a single compact instrument that shares the same imaging optics and line scan camera for both OCT and LSLO imaging. Co-registered high contrast wide-field en face retinal LSLO and SDOCT images are obtained non-mydriatically with less than 600 microwatts of broadband illumination at 15 frames/sec. The LSLO/SDOCT hybrid instrument could have important applications in clinical ophthalmic diagnostics and emergency medicine.

sr-1287



   
 