Abstract: Tracking Scanning Laser Ophthalmoscope (TSLO)

Daniel X. Hammer, R. Daniel Ferguson, John C. Magill, Michael A. White, Ann E. Elsner, "Tracking Scanning Laser Ophthalmoscope (TSLO)," 4951-33, presented at Ophthalmic Technologies XIII (San Jose, CA) , SPIE Proceedings 4951 (January2003).

This paper was published in Ophthalmic Technologies XIII (Proceedings of SPIE, 4951) and is made available as an electronic reprint with permission of SPIE. Single print or electronic copies for personal use only are allowed. Systematic or multiple reproduction, distribution to multiple locations though an electronic listserver or other electronic means, duplication of any material in this paper for a fee or for commercial purposes, or modification of the content of the paper are all prohibited. By choosing to view or print this document, you agree to all the provisions of the copyright law protecting it.

Abstract

The effectiveness of image stabilization with a retinal tracker in a multi-function, compact scanning laser ophthalmoscope (TSLO) was demonstrated in initial human subject tests. The retinal tracking system uses a confocal reflectometer with a closed loop optical servo system to lock onto features in the fundus. The system is modular to allow configuration for many research and clinical applications, including hyperspectral imag- ing, multifocal electroretinography (MFERG), perimetry, quantification of macular and photo-pigmentation, imaging of neovascularization and other subretinal structures (drusen, hyper-, and hypo-pigmentation), and endogenous fluorescence imaging. Optical hardware features include dual wavelength imaging and detection, integrated monochromator, higher-order motion control, and a stimulus source. The system software consists of a real-time feedback control algorithm and a user interface. Software enhancements include automatic bias correction, asymmetric feature tracking, image averaging, automatic track re-lock, and acquisition and logging of uncompressed images and video files. Normal adult subjects were tested without mydriasis to optimize the tracking instrumentation and to characterize imaging performance. The retinal tracking system achieves a bandwidth of greater than 1 kHz, which permits tracking at rates that greatly exceed the maximum rate of motion of the human eye. The TSLO stabilized images in all test subjects during ordinary saccades up to 500 deg/s with an inter-frame accuracy better than 0.05 deg. Feature lock was maintained for minutes despite subject eye blinking. Successful frame averaging allowed image acquisition with decreased noise in low-light applications. The retinal tracking system significantly enhances the imaging capabilities of the scanning laser ophthalmoscope.

Keywords

ophthalmoscopy, confocal imaging, retinal tracking, image stabilization,

sr-1128